Laboratory reaction kettle daily care operation precautions

July 04, 2022

1. In the reaction of different media in the reaction kettle, first check whether the medium has corrosion to the main material. The reaction to the moment is intense, a large amount of gas or a high-temperature flammable and explosive chemical reaction, and the reaction of the high-pressure, ultra-high temperature or the medium containing chlorine ions and fluoride ions to the stainless steel are severely ordered!

2. When loading the reaction medium, it should not exceed 2/3 of the surface of the Reactor body of the laboratory reactor!

3. It is strictly forbidden to dismantle with pressure when working or at the end! It is strictly forbidden to work under overpressure and over temperature!

4. When cleaning the autoclave, special care should be taken not to allow water or other liquid to flow into the heating furnace to prevent the heating furnace wire from being blown. If heating the jacketed heat transfer oil, be careful not to mix water or other liquids when adding the heat transfer oil. Check the oil level of the heat transfer oil from time to time!

5. If there is abnormal sound inside the isolation sleeve during operation, stop the pressure and check whether the mixing system has abnormal conditions. Regularly check the amount of swing of the stirring shaft. If the amount of swing is too large, replace the bearing or sliding bushing in time!

6. According to the voltage of the controller, connect the corresponding voltage, single-phase 220V, three-phase 380V, the fire and zero lines of the controller should not be connected; do not implement wiring work in the power transmission to prevent electric shock! In order to ensure the controller For normal operation and personal safety of the staff, please connect the grounding wire! The control device itself is not explosion-proof, so avoid the inflammable and explosive environment such as oil and gas!

7. When the final reaction temperature of zui is determined, it is not allowed to change the temperature setting value during the heating process, so as to avoid a large overshoot of the temperature!

8. Under the premise of turning on the power, even if the switch such as heating and stirring is not turned on, the output terminal of the rear panel of the controller is also charged. It is not allowed to pull any plug during operation!

9. The working time display is the internal single-chip 8031. It is timed by accumulating the internal partial frequency. Each time the power is turned off, the internal accumulated data will disappear. In the operation, it should be noted that the working time display is divided into units. Working time is 9999 points (166.5 hours)!

10, the speed display correct detection position is essential for the stable display of the speed, this position has been adjusted at the factory, do not adjust at will!

11. Every time you turn on the machine, you should ask any button to be in the initial state. This requires the operator to develop a good working habit. After each work, turn the knob back to the small position of Zui to prevent the current from overshooting when the machine is turned on. Great damage to the control unit!

12. Regularly test various instruments and blasting venting devices to ensure their accurate and reliable work. The working environment of the equipment should meet the requirements of safety technical specifications!

13. When the laboratory reactor is installed, connect the blasting discharge port to the outside through the pipeline!

14. When the laboratory reactor is deactivated for a long time, the inside and outside of the kettle should be cleaned and cleaned without water and other materials, and stored in a clean, dry and non-corrosive area.

The above is the Laboratory reaction kettle daily care operation precautions we have listed for you. You can submit the following form to obtain more industry information we provide for you.

You can visit our website or contact us, and we will provide the latest consultation and solutions

Share to:

Send Inquiry

Home> Industry Information> Laboratory reaction kettle daily care operation precautions
Related Products List

Home

Product

Phone

About Us

Inquiry

We will contact you immediately

Fill in more information so that we can get in touch with you faster

Privacy statement: Your privacy is very important to Us. Our company promises not to disclose your personal information to any external company with out your explicit permission.

Send